PRODUCTS NAME
detectability
Detection capability
1SEC
Detection speed
Detection accuracy
Repeatability accuracy
GLOBAL PARTNER
DEMO VIDEO
Intraocular Lens Compression Force Tester is specifically designed to measure the compression force when intraocular lens haptics are compressed to the nominal diameters of 10 mm or 11 mm, fcomplying with ISO standard
Compliant with ISO 11979-3
Electric thermostatic bath delivers stable solution test environment
Supplied with 10 mm and 11 mm standard weight (already calibrated)
Optional camera system can conduct real-time monitoring of the IOL compression process
Two clampers with surface radii of 5.00mm and 5.50mm (accuacy within 0.02mm).
Low-friction material shall be used to minimize rotational resistance of haptics.
Force gauge accuracy up to 0.01mN
Measurement Step
1.Place the device horizontally to avoid impact from gravity
2.Adjust the clamper size to be the same size as IOL and then put IOL onto it
3.Tighten the clamper to the norminal dimater of IOL
4. Read compression force value on the LCD sreen

(First step)
Sample testing process